Programat P95 Manual: Ivoclar

If a third-party pyrometer shows the P95 is firing too hot/cold, the operator can apply an offset (e.g., -5°C) without adjusting every program. Manual caution: Do not exceed ±10°C offset.

Best practice: Download the latest PDF from Ivoclar and print the “Quick Reference” pages (usually pages 10–15 and the troubleshooting flowchart). Laminate those pages and hang them near the unit. Ivoclar Programat P95 Manual

The P95 uses a color-coded, icon-driven touchscreen. The manual provides a detailed map of: If a third-party pyrometer shows the P95 is

Without the manual, technicians often miss the "Calibration" submenu, which is essential for maintaining accuracy. Without the manual, technicians often miss the "Calibration"

If restorations come out over-fired (too glossy) or under-fired (matte, chipping), run a calibration test. The manual outlines using the "Calib" program with a ceramic calibration pellet. The target temperature difference should be less than 5°C. If higher, adjust the "Temp. Offset" parameter in the service menu (accessible with code 1234).